Triggering - The TRIGger Subsystem
Measurement Operations & Settings
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
179
Extended Trigger Configuration
This section includes information for advanced users about how to 
customize your use of the trigger system. 
You can configure the ouputs and inputs from two nodes, Node A and 
Node B. See Figure 7  on page 180 for more information on Node A and 
Node B. You can configure these nodes to be triggered by certain events 
and for these nodes to trigger particular actions.
command:
:TRIGger
syntax:
:TRIGger<wsp>NODEA|1|NODEB|2
description:
Generates a hardware trigger.
parameters:
1 or NODEA:
2 or NODEB:
Generates trigger at Node A. 
Generates trigger at Node B.
N O T E
“:TRIGger” on page 172 describes the :TRIGger command for basic users.
response:
none
example:
trig 1
command:
:TRIGger:CONFiguration:EXTended
syntax:
:TRIGger:CONFiguration:EXTended<wsp><Node A Input Config.>,
<Node B Input Config.>,<Output Matrix Config.>
description:
Sets the extended hardware trigger configuration. 
parameters:
Node A Input Configuration:
Node B Input Configuration:
Output Matrix Configuration:
32-bit unsigned integer, see below.
32-bit unsigned integer, see below.
32-bit unsigned integer, see below.
response:
none
example:
trig:conf:ext 0,0,0
command:
:TRIGger:CONFiguration:EXTended?
syntax:
:TRIGger:CONFiguration:EXTended?
description:
Returns the extended hardware trigger configuration.
parameters:
none
response:
Node A Input Configuration:
Node B Input Configuration:
Output Matrix Configuration:
32-bit unsigned integer, see below.
32-bit unsigned integer, see below.
32-bit unsigned integer, see below.
example:
trig:conf:ext? 
→ +0,+0,+0<END>