Introduction to Programming
The Status Model
36
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
Status System for 8166A/B
The status system for the Agilent 8166A/B Lightwave Multichannel 
System returns the status of 17 module slots. The 
Operational/Questionable Status Summary Registers consists of two 
levels, as described by Figure 5 
Module slots 1 to 14 affect the Level 0 summary register as described in 
Figure 4 . Bit 0 of the Level 0 summary registers represents the summary 
of the status of module slots 15, 16, and 17. The Level 1 summary registers 
contain an individual summary for each of these module slots.
Figure 5 The Operational/Questionable Status System for 8166A/B
Operational/Questionable Status
Operational/Questionable  Status
Operational/Questionable  Status
Condition Summary Register (Level 0) 
Event Summary Register (Level  0)
Enable Summary Mask 
Register (Level 0)
Status Byte
Status Byte Register
Status 
Summary
to next 
&
&
&
&
OR
level
to next 
Operational/Questionable 
Operational/Questionable 
Operational/Questionable
Slot Status Condition 
Slot Status Event
Slot Status Enable Mask 
&
&
&
&
OR
level
to next 
for a positive
&
&
&
&
OR
level
Status
Summary
Slot 15
transition
for a positive
transition
for a positive
transition
Operational/Questionable Status  
Operational/Questionable  Status
Operational/Questionable  Status
Condition Summary Register (Level 1) 
Event Summary Register (Level  1)
Enable Summary Mask
Register
Register 
Register 
Register (Level 1)
for Level 1
for Level 0