*STB?
returns status byte, value 0 .. +255
*ESE
sets the standard event status enable mask, parameter 0 .. +255
*ESE?
returns SESE, value 0 .. +255
*ESR?
returns the standard event status register, value 0 .. +255
*OPC
parses all program message units in the message queue, and prevents 
the instrument from executing any further commands until all pending 
commands are completed.
*OPC?
returns 1 if all operations (scan trace printout, measurement) are com
pleted. Otherwise it returns 0.
*CLS
clears the status byte and SESR, and removes any entries from the er
ror queue.
*RST
clears the error queue, loads the default setting, and restarts communi
cation. 
NOTE: *RST does NOT touch the STB or SESR. A running measurement 
is stopped.
*TST?
initiates an instrument selftest and returns the results as a 32 bit 