Specific Commands
Specific Command Summary
44
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
Specific Command Summary
The commands are ordered in a command tree. Every command belongs to
a node in this tree.
a node in this tree.
The root nodes are also called the subsystems. A subsystem contains all
commands belonging to a specific topic. In a subsystem there may be
further subnodes.
commands belonging to a specific topic. In a subsystem there may be
further subnodes.
All the nodes have to be given with a command. For example in the
command
command
disp:brig
•
DISPlay
is the subsystem containing all commands for controlling the
display,
•
BRIGhtness
is the command selecting brightness.
Table 4 gives an overview of the command tree. You see the nodes, the
subnodes, and the included commands.
subnodes, and the included commands.
If a command and a query are both available, the command ends
/?
.
So,
disp:brig/?
means that
disp:brig
and
disp:brig?
are both available.
N O T E
Table 4 Specific Command Summary
Command
Description
Page
:CONFigure[n][:CHANnel[m]]:OFFSet
:WAVelength:REFerence/?
Sets or queries the slot and channel of the external reference
powermeter.
powermeter.
:WAVelength:STATe/?
Switches or queries attenuator Offset Table on or off/?
:WAVelength:TABle?
Queries the complete offset table.
:WAVelength:TABle:SIZE?
Queries the size of the offset table.
:WAVelength:VALue
Adds a value pair (wavelength, offset) to the offset table.
:WAVelength:VALue:DELete
Deletes an offset value pair.
:WAVelength:VALue:DELete:ALL
Deletes all value pairs from the offset table.
:WAVelength:VALue:OFFSet?
Queries an offset value according to wavelength or index.
:WAVelength:VALue:PAIR?
Queries an offset/wavelength value pair according to wavelength
or index.
or index.
:WAVelength:VALue:WAVelength?
Queries a wavelength value from its index in the offset table