Instrument Setup and Status
Status Reporting – The STATus Subsystem
64
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
Status Reporting – The STATus 
Subsystem
The Status subsystem allows you to return and set details from the Status 
Model. For more details, see “The Status Model” on page 33.
command:
:STATus:OPERation[:EVENt][:LEVel0]?
syntax:
:STATus:OPERation[:EVENt][:LEVel0]?
description:
Returns the Operational Status Event Summary Register (OSESR).
parameters:
none
response:
The sum of the results for the slots (a 16-bit signed integer value, where 0 
≤ value ≤ 32767):
Bits
Mnemonics
Decimal Value
Agilent 8163A/B
Agilent 8164A/B
Agilent 8166A/B
15
Not used
Not used
Not used
0
14
Not used
Not used
Slot 14 Summary
16384
13
Not used
Not used
Slot 13 Summary
8192
12
Not used
Not used
Slot 12 Summary
4096
11
Not used
Not used
Slot 11 Summary
2048
10
Not used
Not used
Slot 10 Summary
1024
9
Not used
Not used
Slot 9 Summary
512
8
Not used
Not used
Slot 8 Summary 
256
7
Not used
Not used
Slot 7 Summary
128
6
Not used
Not used
Slot 6 Summary
64
5
Not used
Not used
Slot 5 Summary
32
4
Not used
S
lot 4 S ummary 
S
lot 4 S ummary 
16
3
Not used
Slot 3 Summary
Slot 3 Summary
8
2
Slot 2 Summary
Slot 2 Summary
Slot 2 Summary
4
1
Slot 1 Summary
Slot 1 Summary
Slot 1 Summary
2
0
Not used
Slot 0 Summary
Level 1 Summary 
1
example:
stat:oper? 