Appliance dimensions
Adjustable-height feet
Description of the appliance
Type plate
Operating and control 
elements
Opening for external tem-
perature sensor
Other features
-  Audible and visual temperature alarm (adjustable limits).
-  Audible and visual door open alarm.
-  Floating contact for connection to a remote monitoring system.
-  Serial interface (RS485) for external temperature and alarm 
documentation.
-  Maximum/minimum interior temperatures are stored.
-  Last 3 temperature alarms are saved with time, date and duration 
of alarm.
-  Last 3 power cuts are saved with time, date and duration of power 
cut.
-  Opening for installing a reference sensor.
-  Safety thermostat to avoid temperatures below +2°C.
It is essential to use these safety facilities to avoid damage 
to stored items. These facilities must not be deactivated or 
decommissioned!
Range of appliance use
The appliance is suitable for storing and cooling of 
laboratory preparations at temperatures of between 
+3°C and +8°C.
The appliance is not suitable for use in explosion-hazard areas.
For the storage of valuable or temperature-sensitive substances or 
products the use of an independent, constantly monitoring alarm 
system is necessary.
This alarm system must be designed so that each alarm status is 
detected immediately by an authorised person who can then take 
appropriate action.
Position the temperature sensor for this system in the upper part 
of the interior container (see Opening for external temperature 
sensor).
Disposal notes
The appliance contains reusable materials and should 
be disposed of properly - not simply with unsorted 
household refuse. Appliances which are no longer 
needed must be disposed of in a professional and 
appropriate way, in accordance with the current local 
regulations and laws. 
When disposing of the appliance, ensure that the refrigerant circuit 
is not damaged to prevent uncontrolled escape of the refrigerant 
it contains (data on type plate) and oil. 
•  Disable the appliance.
•  Pull out the mains plug.
•  Cut through the connection cable.
 WARNING
Danger of suffocation due to packing material and plastic film!
Do not allow children to play with packaging material. Take the
packaging material to an official collection point.
Interior light (in appli-
ances with glass door)
Climate rating
The climate rating indicates the room tem-
perature at which the appliance may be 
operated in order to achieve full refrigeration 
performance.
The climate rating is indicated on the type 
plate.
The position of the type plate is shown in the section entitled 
Description of the appliance.
Climate rating 
Room temperature
SN 
+10°C to +32°C
+16°C to +32°C
ST 
+16°C to +38°C
+16°C to +43°C
SN-ST 
+10°C to +38°C
SN-T 
+10°C to +43°C
Do not operate the appliance outside the speci-