Helio PM5500 Manual De Usuario

Descargar
Página de 16
13
H
ELI
O
2
BLENDEr
HELIUM-OxygEn
TROUBLESHOOTING
If the
 
H
ELI
O
2
 
Blender fails to function, consult the Troubleshooting 
Guide below.
If problem cannot be solved by using Troubleshooting Guide, refer to the
 
H
ELI
O
2
 
Blender Service Manual (P/N 506124) available on the Internet; 
www.precisionmedical.com or consult your Provider.
Problem
Probable Cause
Remedy
Oxygen 
concentration 
discrepancy 
between Blender 
setting and 
alarmed Oxygen 
Monitor / Analyzer 
(greater than 3%)
1. •HIgH flow model, 
flow requirement 
below 15 lpm.
LOW flow model, 
flow requirement 
below 3 lpm.
2. Alarmed Oxygen 
Monitor / Analyzer 
inaccurate
3. Low flow bleed 
obstructed
4. Gas supply 
contaminated or 
heliox concentration 
incorrect
5. Downstream device 
causing back flow 
or restricted flow
6. Supply pressure 
imbalanced
1. Use auxiliary outlet & 
engage bleed
2. Recalibrate alarmed 
Oxygen Monitor / 
Analyzer or Verify 
with second alarmed 
Oxygen Monitor / 
Analyzer
3. Remove obstruction
4. Check gas sources 
with calibrated alarmed 
Oxygen Monitor / 
Analyzer to confirm 
oxygen is 100% and 
verify heliox tank content
5. Isolate Blender. Check 
oxygen concentration at 
Blender Outlets
6. Assure heliox and 
oxygen inlets pressures 
are within 10 psi
No flow at
Blender outlets
1. Gas sources turned 
“OFF”
2. Gas sources not 
connected
1. Turn gas sources “ON”
2. Connect gas sources
Alarm sounding
1. Difference between 
oxygen and heliox 
inlet pressures 
greater than specified
1. Correct pressure 
difference until heliox 
and oxygen pressures 
are within specification