HELIUM-OxygEn
WARNING
•  Only trained, qualified medical personnel under the direct supervision of a 
licensed physician should operate the 
H
ELI
O
2
 Blender.
•  Use this 
H
ELI
O
2
 Blender only for its Intended Use as described in this 
manual.
•  Confirm prescribed dose before administering to patient. Monitor on a 
frequent basis. Continuous Monitoring with an alarmed Oxygen Monitor 
/ Analyzer is recommended.
•  The 
H
ELI
O
2
 Blender shall be serviced by a qualified service technician.
•  Always follow ANSI and CGA standards for Medical Gas Products, 
Flowmeters and Oxygen Handling. 
•  The clinician must verify the Heliox tank concentration prior to utilizing 
the Blender.
•  An alarmed Oxygen Monitor / Analyzer must be used to verify oxygen 
concentration.
•  Accuracy of oxygen concentration will be affected if bleed is not activated 
at flow settings below 15 lpm for the High Flow Blender, and 3 lpm for 
the Low Flow Blender.
•  dO NOT obstruct the alarm.
•  dO NOT use Blender when alarm is sounding.
•  dO NOT use oil in or around the Blender.
•  dO NOT occlude or obstruct the bleed port on the auxiliary outlet of 
the Blender.
•  dO NOT use near any type of flame or flammable/explosive 
substances, vapors or atmosphere.
•  Oxygen Concentration dial does not rotate 360 degrees. Rotating the 
dial beyond the endpoint settings will damage the Blender.
CAUTION
•  Turn off gas supplies when
H
ELI
O
2
 Blender is not in use.
•  Store the
H
ELI
O
2
 Blender in a clean, dry area when not in use.
•  The
H
ELI
O
2
 Blender contains magnetic, ferrous material that may
affect the results of an MRI.
•  Ensure all connections are tight and leak free.
•  Avoid excessive pressure surges greater than 100 psi (6.9 bar) when
pressuring the Blender inlets.
•  DO NOT steam autoclave.
•  DO NOT immerse
H
ELI
O
2
 Blender into any liquid.
•  DO NOT gas sterilize with (EtO) Ethylene Oxide.
•  DO NOT use if dirt or contaminants are present on or around the
Blender or connecting devices.
•  DO NOT smoke in an area where oxygen is being administered.
•  DO NOT clean with aromatic hydrocarbons.
•  Inlet pressure of Device used in conjunction with the
H
ELI
O
Blender must
match inlet pressure of
F
I
O
2
 
Blender.