PROGRAMMING FUNCTIONS - LEVEL  2 (cont.)
SERVICE TOOLS
This function allows the testing of individual components 
and the ability to check switches for proper function. This 
function also tests the funnel sensor coil’s frequency 
(diagnostic tool for troubleshooting purposes only).  
Procedure to test components and outputs:
The following components can be individually tested:
Left Brew Valve
Left Bypass Valve
Left Funnel Lock
Right Brew Valve
Right Bypass Valve
Right Funnel Lock
Refill Valve
Tank Heater Relay
Tank Heater Triac
1.  Place a funnel into the rails on both sides of the 
brewer.
2.  Place a server beneath each funnel.
3.  Press and hold the right hidden button until the dis-
play reads SET LANGUAGE. Press the right hidden 
button until the display reads SERVICE TOOLS.
4.  Press  and  release  YES.  The  display  should  now 
read TEST OUTPUTS? Press and release YES to 
test individual components and outputs. Pressing 
NO will advance to the next programming function, 
TEST SWITCHES.
5.  The display should now read <- <- BREW VALVE.  
To test the left brew valve, press ON. If the brew 
valve is functional, water should run from the left 
funnel.
6.  Press OFF to end the flow of water.
7.  Press NEXT to advance to the next component to 
be tested.
NOTE: To bypass testing any component, press NEXT 
to advance to the next component without testing the 
previous one.
8.  The display should now read <- <- BYPASS VALVE
To test the left bypass valve, press ON. If the bypass 
valve is functional, water should run from the left 
funnel.
9.  Press OFF to end the flow of water.
10. Press NEXT to advance to the next component to 
be tested, <- <- FUNNEL LOCK.
11. To  test  the  left  FUNNEL  LOCK,  press  ON.  If  the 
funnel lock is functional, the lock will drop to hold 
funnel in place.
12. Press OFF to retract the funnel lock.
13. Follow  steps  5-12  to  test  the  right  side  compo-
nents.
14. To test the REFILL VALVE, press ON. If the refill 
valve is functional, the sound of the valve operating 
will be heard.
15. Press OFF to end testing of the refill valve.
16. Press NEXT to advance to the next component to 
be tested, TANK HEATR RELAY.
17. To test the tank heater relay, connect a voltmeter 
across each of the tank heaters to check for volt-
age.
18. Press ON. The correct voltage should be present 
at the heater terminals.
19. Press OFF to end testing of the tank heater relay.
NOTE: The tank heater will automatically turn off if left 
on too long.
20. Press NEXT to advance to the next component to 
be tested, TANK HEATR TRIAC.
21. To test the tank heater triac, connect a voltmeter 
across each of the tank heaters to check for volt-
age.
22. Press ON. The correct voltage should be present 
at the heater terminals.
23. Press OFF to end testing of the tank heater triac.
24. Press NEXT to return to the TEST OUTPUTS screen. 
To  advance  to  the  next  function  screen,  TEST 
SWITCHES, press NO. To exit programming, press 
and release either ON/OFF button to return to the 
MAIN SCREEN.
40119 071107