Tektronix AWG710B Manual De Usuario
Appendix B: Performance Verification (AWG710B)
AWG710&AWG710B Arbitrary Waveform Generator User Manual
B-35
Figure B-16: Trigger signal and waveform output (+5 V check 1)
e.
Change the pulse waveform high level to 5.35 V by using cursor keys.
f.
Check that sine waves are displayed on the oscilloscope as shown in Figure
B-7.
B-7.
Figure B-17: Trigger signal and waveform output (+5 V check 2)
7. Follow the substeps below to check trigger level accuracy at –5 V:
a. Push –, 5 and ENTER keys in this order.
b. Set the DC waveform low level of function generator to –5.35 V by using
cursor keys.
Trigger
signal
Waveform
output
CH1
CH2
Trigger
signal
Waveform
output
CH1
CH2