Clarity 87392-11 Manual De Usuario

Descargar
Página de 53
Clarity User Guide,
 
March 1, 2011 - Rev. 1.06 
                    
           Page 11
 
Figure 1.5 – Installing the Glass Plate 
 
 
Installing the Glass Plate – The glass plate is placed over the opening in the base plate of the 
blocking chamber.  Clean the glass plate and then slide it into the pocket.  It should fit and 
seat down in the pocket fully.  (The glass plate is shipped in the accessory box.) 
 
Installing Light Shielding – This shielding is designed to reduce or eliminate light coming in 
to the blocking chamber causing inconsistent lens analysis.  The shields are not necessary in 
all lab environments.  Only if you see white or clear areas on the lens image (on screen) 
during processing should the shields be needed.  They are in the accessory box in the event 
that there is too much light entering the chamber.  There are three separate parts to the light 
shielding including the left gripper shield, the right gripper shield and the blocking arm shield. 
 
The left and right gripper shields can be installed using the screws in the accessory box to 
secure them to the tops of the front gripper arms.  There are two threaded holes for the screws 
to attach the shields to the tops of each gripper arm. 
 
The blocking arm shield requires that you remove the center screw of the blocking arm, 
thereby removing the chuck.  Place the shield over the arm with the flat side facing down.  
Replace the chuck (align small hole on the back of the chuck with the locator pin) and insert 
and tighten the center screw.