National Instruments NI 5422 Manual De Usuario

Descargar
Página de 36
© National Instruments Corporation
3
NI 5422 Specifications
Amplitude and Offset
Amplitude 
Range
Path
Load
Amplitude (V
pk-pk
)
1. Amplitude 
values assume 
the full scale 
of the DAC is 
utilized. If an 
amplitude 
smaller than 
the minimum 
value is 
desired, then 
waveforms less 
than full scale 
of the DAC can 
be used.
2. NI-FGEN 
compensates 
for user-
specified 
resistive loads.
Minimum Value
Maximum Value
Direct
50 
Ω
0.707
1.00
1 k
Ω
1.35
1.91
Open
1.41
2.00
Low-
Gain 
Amplifier
50 
Ω
0.00564
2.00
1 k
Ω
0.0107
3.81
Open
0.0113
4.00
High-
Gain 
Amplifier
50 
Ω
0.0338
12.0
1 k
Ω
0.0644
22.9
Open
0.0676
24.0
Amplitude 
Resolution
3 digits 
Offset Range
Span of ±50% of Amplitude Range with increments 
<0.0028% of Amplitude Range.
Not available on 
the Direct Path.
Table 1.   (Continued)
Specification
Value
Comments