National Instruments NI 5422 Manual De Usuario

Descargar
Página de 36
© National Instruments Corporation
9
NI 5422 Specifications
Table 1.  (Continued)
Specification
Value
Comments
Spectral Characteristics
Spurious-Free 
Dynamic 
Range 
(SFDR) with 
Harmonics
Path
Amplitude 
–1 dBFS.
Measured from 
DC to 100 MHz.
Also called 
harmonic 
distortion.
SFDR with 
harmonics at low 
amplitudes is 
limited by a 
–148 dBm/Hz 
noise floor. All 
values are 
typical and 
include aliased 
harmonics.
Direct
Low-Gain 
Amplifier
High-Gain 
Amplifier
1 MHz
–70 dBc
–65 dBc
–66 dBc
5 MHz
–70 dBc
–65 dBc
–58 dBc
10 MHz
–70 dBc
–65 dBc
–52 dBc
20 MHz
–63 dBc
–64 dBc
–49 dBc
30 MHz
–57 dBc
–60 dBc
–43 dBc
40 MHz
–48 dBc
–53 dBc
–39 dBc
50 MHz
–48 dBc
–53 dBc
60 MHz
–47 dBc
–52 dBc
70 MHz
–47 dBc
–52 dBc
80 MHz
–41 dBc
–52 dBc