Bunn ICB Guía Del Usuario

Descargar
Página de 70
Page 48
PROGRAMMING THE BREWER (cont.)
PROGRAMMING FUNCTIONS - LEVEL  2 (cont.)
SERVICE TOOLS
This function allows the testing of individual components 
and the ability to check switches for proper function. This 
function also tests the funnel sensor coil’s frequency 
(diagnostic tool for troubleshooting purposes only).  
Procedure to test components and outputs:
The following components can be individually tested:
Brew Valve
Bypass Valve
Refi ll Valve
Tank Heater Relay
Tank Heater Triac
Funnel Lock (Optional)
1.  Place a funnel into the rails of the brewer.
2.  Place a server beneath the funnel.
3.  Press and hold the right hidden button until the dis-
play reads SET LANGUAGE. Press the right hidden 
button until the display reads SERVICE TOOLS.
4. Press and release YES. The display should now 
read  TEST OUTPUTS? Press and release YES to 
test individual components and outputs. Pressing 
NO will advance to the next programming function, 
TEST SWITCHES.
5.  The display should now read BREW VALVE.  To 
test the brew valve, press ON. If the brew valve is 
functional, water should run from the funnel.
6. Press 
OFF to end the fl ow of water.
7. Press 
NEXT to advance to the next component to 
be tested.
NOTE: To bypass testing any component, press NEXT 
to advance to the next component without testing the 
previous one.
8.  The display should now read BYPASS VALVE. To 
test the bypass valve, press ON. If the bypass valve 
is functional, water should run from the funnel.
9. Press 
OFF to end the fl ow of water.
10. Press NEXT to advance to the next component to 
be tested, REFILL VALVE.
11. To test the REFILL VALVE, press ON. If the refi ll 
valve is functional, the sound of the valve operating 
will be heard.
12. Press OFF to end testing of the refi ll valve.
13. Press NEXT to advance to the next component to 
be tested, TANK HEATR RELAY.
14. To test the tank heater relay, connect a voltmeter 
across each of the tank heaters to check for volt-
age.
15. Press ON.  The correct voltage should be present 
at the heater terminals.
16. Press OFF to end testing of the tank heater relay.
NOTE: The tank heater will automatically turn off if left 
on too long.
17. Press NEXT to advance to the next component to 
be tested, TANK HEATR TRIAC.
18. To test the tank heater triac, connect a voltmeter 
across each of the tank heaters to check for volt-
age.
19. Press ON.  The correct voltage should be present 
at the heater terminals.
20. Press OFF to end testing of the tank heater triac.
21. Press NEXT to advance to the next component to 
be tested, FUNNEL LOCK.
22. To test the FUNNEL LOCK, press ON. If the funnel 
lock is functional, the lock will drop to hold the 
funnel in place.
23. Press OFF to retract the funnel lock.
24. Press NEXT to TEST OUTPUTS?. To exit program-
ming, press and release the ON/OFF button to return 
to the MAIN SCREEN.
37077 101205