8.2.4 Exciting Frequency 
 
Proceed as follows to check the Exciting Frequency 
Switch operation 
Display example 
Description 
 
A:DETECTOR 
1 EXC CUR 
2 SIZE 
3 EXC FREQ
 
4 FLOW DIR 
5 EXIT 
 
Select "EXC FREQ" from the setting item 
selection menu. 
 
 
STEP1 
 
 
 
EXC FREQ 
 
200 
 
       Hz 
 
 
The currently set excitation frequency 
(200Hz in this example) appears. 
 
Then push              . 
 
* Pushing        returns you to the setting 
menu. 
 
 
 
8.2.5 Flow Direction Setting 
 
In the converter, you can set the flow direction of fluid arbitrarily. 
 
● 
Flow direction setting 
 
Selection item 
Contents 
NORMAL 
When the fluid flows in the direction of the arrow 
indicating the flow direction that is attached to the 
detector, the indicator value and electric current 
output value increase. 
SWITCH 
When the fluid flows in the reverse direction of the 
arrow indicating the flow direction that is attached to 
the detector, the indicator value and electric current 
output value increases. 
 
Switch operation 
Display example 
Description 
 
A:DETECTOR 
1 EXC CUR 
2 SIZE 
3 EXC FREQ 
4 FLOW DIR
 
5 EXIT 
 
Select "FLOW DIR" from the setting item 
selection menu. 
 
 
 
 