 83 − 
 
6 F 8 A 0 9 3 4  
 
●Digital input control signal 
You can select the detective level of the digital input, as shown below, to control the totalizer 
and pulse output.   
(Only when the digital input function is set for totalizer control input) 
Selection items 
Digital input function setting
Totalizer control signal 
CNT ST/SP 
(Totalizer START/STOP)
H signal:Totalizer STOP 
L signal:Totalizer START 
L LEVEL 
CNT RS/ST 
(Totalizer RESET/START)
H signal:Totalizer START 
L signal:Totalizer RESET 
CNT ST/SP 
(Totalizer START/STOP)
H signal:Totalizer START 
L signal:Totalizer STOP 
H LEVEL 
CNT RS/ST 
(Totalizer RESET/START)
H signal:Totalizer RESET 
L signal:Totalizer START 
 
●Changing the digital input function 
The following is an example to change the Digital Input (DI) function from No use (NO 
USE
) to    Totalizer Start / Stop (CNT ST/SP). 
Switch operation 
Display example 
Description 
 
I:DI 
1 DI FUNC
 
2 DET LVL 
3 EXIT 
 
 
 
 
Select "DI FUNC" from the setting item 
selection menu. 
STEP1 
 
 
 
DI FUNC 
 
NO USE 
 
 
 
 
 
The current setting (NO USE in this 
example) appears. 
 
Then push              . 
 
* Pushing       returns you to the menu 
screen. 
STEP2 
 
 
 
DI FUNC 
 
NO USE
 
 
        
 
 
The switches at the bottom change. 
(                                      are shown.) 
At the same time, the cursor appears. 
(The item indicated by the cursor is 
highlighted.) 
 
Then push                          . 
STEP3 
 
 
 
DI FUNC 
 
CNT ST/SP
 
 
        
 
 
You can continue to change the setting item.
Push                          to  change  the  selection 
items. 
: Selected item is scrolled up. 
: Selected item is scrolled down. 
When the desired item is selected, push         
    to set the item temporarily. 
 
ESC 
ESC