Toshiba scale view v4 Manual De Usuario
Error Messages
Scale View V4.0 Owner’s Manual
Page 32
18-Oct-04
“Run Time error 76”
when trying to access
TMCC log
when trying to access
TMCC log
The file “notepad.exe” is not found in the default
directory. The default directory for Scale View for
“notepad.exe” is “C:\Windows”. This error usually
occurs with Windows NT, Windows NT WS and
Windows NT 2000.
directory. The default directory for Scale View for
“notepad.exe” is “C:\Windows”. This error usually
occurs with Windows NT, Windows NT WS and
Windows NT 2000.
To correct this first confirm that there is not a
“Windows” subdirectory. If there is no “Windows”
subdirectory you must create one and copy the file
“notepad.exe” into it. This should take care of this
error.
“Windows” subdirectory. If there is no “Windows”
subdirectory you must create one and copy the file
“notepad.exe” into it. This should take care of this
error.
An error occurs when
requesting a report
requesting a report
“Unable to connect, incorrect session parameters.”
Or Microsoft Access opens and requests a
password.
Or Microsoft Access opens and requests a
password.
These problems occur because of a conflict between
different versions of Crystal Reports installed by
other programs.
different versions of Crystal Reports installed by
other programs.
Contact your TEC Representative for assistance.
Scale Not
Communicating
Communicating
In Scale View V4.0 this error indicates that there is a
problem with a Scale’s IP address. This problem
could be that it isn’t programmed in the scale or in
Scale View’s Scale Table correctly. Check the Scale
View Scale Table configuration and the Scale’s IP
address.
problem with a Scale’s IP address. This problem
could be that it isn’t programmed in the scale or in
Scale View’s Scale Table correctly. Check the Scale
View Scale Table configuration and the Scale’s IP
address.