8.0  CALIBRATION 
 
The GBC/GFC scales are calibrated using metric weights and GBC/GFC-a 
scales  are  calibrated  using  metric  or  pound  weights  depending  on  the 
unit  in  use  before  calibration.    See  the  Service  parameters  section  for 
more information. 
 
The  scale  will  display  a  value  of  the  weight  to  be  used  for  calibration, 
this  value  is  the  last  weight  used  for  calibration.    You  can  enter  a 
different value, if desired. 
 
For entering Calibration, see section 7.0 Parameters 
 
When “F1”  “CAL” is displayed, press [Tare]
 
The display will then show "unLoAd" to request all weight be  removed 
from the platform.    
 
Press the [Tare] key to set the zero point. 
 
The displays will then show the calibration weight suggested, the value 
is shown as an integer number, it is not possible to have fractions of a 
kilogram or pound.   For example:    
LoAd 
00004 
KiLoS 
 
Place the calibration weight on the platform. If the calibration weight is 
different  from  the  value  shown,  Press  [CE]  to  clear  the  current  value 
then enter the correct value.    
 
Press [Tare] to calibrate. 
 
When calibration is done the display will show “SPAn”  “PASS”. 
 
Remove the weight. The scale will return to normal weighing.   
 
If an error message “SPAn” “FAIL” is displayed, then repeat the process. 
 
After calibration, the scale should be checked whether the calibration is 
correct. If necessary, repeat calibration. 
NOTE: GBC/GFC-a scales will have the lb or kg indicator on, to show the unit of the 
weight  requested.  If  the  scale  was  in  pounds  before  starting  the  calibration,  the 
weights requested will be in pound values or if the scale was weighing in kilograms 
then metric weights will be requested.