fisher--paykel mr850 Manual De Usuario

Descargar
Página de 62
 
MR850 Technical Manual   *   Revision J   * Issued November 2005   *   Ref.  185041340 (185041713 internal)
 
17
4.2.3 Mute 
Button 
 
The mute button silences the humidifier's audible alarm.  The muted time 
depends on the alarm condition.  In general, alarms will be muted for 2 minutes. 
A chamber or airway probe alarm is muted for a longer time, until the humidifier 
determines whether the probe is in or out.  The temperature alarm is treated 
differently - see section 4.5. 
4.3 Temperature 
Display 
The front panel shows the lower of the chamber or airway temperatures.  This temperature 
gives an indication of the dew point (in °C) of the gas that is being supplied to the patient.  The 
dew point of a gas is the best indication of both its humidity and energy content.  Under normal 
operation, the displayed temperature will be the chamber temperature, as its control set point is 
lower.  If the temperature is above 70 °C, "Hi" will be displayed.  If the temperature is below 10 
°C, "Lo" will be displayed.  If HC mode has been enabled the decimal point on the temperature 
display will flash. 
4.3.1 
Showing Chamber and Airway Temperature 
Both the chamber and airway temperature can be displayed by pushing and holding the mute 
button for 1 second.  The temperatures are displayed in the following sequence: 
1  Chamber temperature is displayed until two seconds after the mute button is released.  The 
chamber probe indicator (see section 4.4) will also light to show which temperature is being 
displayed. 
2  The display will blank, and then the airway temperature will be displayed until two seconds 
after the mute button is released.  The airway probe indicator will also light (see section 4.4) 
to show which temperature is being displayed. 
3  The temperature display will blank again, and revert to normal operation. 
4.4 Setup 
Indicators 
The MR850 setup indicators, placed on the lower left of the front panel, are intended to aid the 
user in identifying problems with the incorrect setup of the device and its accessories. 
4.4.1 
Heater wire connector 
 
These indicators light if the heater wire in the breathing circuit has not been 
connected correctly, or if the heater wire or heater wire adaptor is faulty.  An 
intermittent connection, or excessive current (total current in all limbs > 3.5 A) in 
the heater wires will also produce this alarm.  The humidifier will remove power 
from the heating systems if this alarm is active. 
NOTE: Software versions 5.45 and 5.70 do not light the upper indicator located 
on the breathing circuit. 
NOTE: Software versions 7.22 and 7.23 in non heater wire operation and 
without a heater wire connected no audible or visual heater wire alarm will be 
given.  Connecting a heated wire circuit will automatically cause the MR850 to 
default to heater wire operation.