MR850 Technical Manual   *   Revision J   * Issued November 2005   *   Ref.  185041340 (185041713 internal)
 
21
5 Maintenance 
Procedures 
In order to keep your humidifier in good working order, it is necessary to perform maintenance 
at regular intervals. 
5.1 Maintenance 
Schedule 
5.1.1 MR850 
Humidifier 
Annually 
a.  Check MR850 for physical damage: 
•  Check the mains cable for damage, replace if necessary. 
•  Check the heater plate for deep scratching etc., replace if necessary. 
•  Check the heater wire adaptor for kinks, abrasions and damaged connectors.  Check 
that the plugs couple with the sockets on the humidifier. 
 
b.  Carry out a full performance test.  This procedure is detailed in section 8. 
 
NOTE
: A maintenance check sheet is given in Section 9. 
 
5.1.2 
MR850 Temperature Probe 
Every Six months 
a.  Visually check the humidifier probes for physical damage: 
•  Check that the chamber probe's glass thermistor has not been damaged.  Replace 
probe if required. 
•  Inspect the chamber probe's glass thermistor for deposits or foreign material.  Clean 
probe as required. 
•  Check the probe cable for kinks and abrasions etc. 
•  Check that the probe connectors couple with the humidifier sockets. 
 
b.  A temperature accuracy check and flow accuracy check should be performed on the MR850 
temperature probe as outlined in section 8.2.