MR850 Technical Manual   *   Revision J   * Issued November 2005   *   Ref.  185041340 (185041713 internal)
 
27
6.3 
“See Manual” Error Codes 
The following is an explanation of the error codes that are displayed in conjunction with the See 
Manual indicator flashing.  A code is not displayed if the microprocessor has stopped 
functioning (see technical problems - section 6.2). 
 
Error  Description of Fault 
E00 No 
fault 
E02 
Microprocessor stack overflow 
E03 RAM 
fault 
E04 ROM 
fault 
E05 
EEPROM version older than ROM version.  Update EEPROM.  Refer to Section 7.2.6 
E06 
EEPROM version newer than ROM version – (old software).  Contact Fisher & Paykel 
Healthcare 
E07 
Model mismatch with software, contact your Fisher & Paykel Healthcare 
representative. 
NOTE: included in software version 7.14 onward, except 7.21. 
E10 
Temperature circuit calibration out of range: Range Amp 0: 25.5 °C 
E11 
Temperature circuit calibration out of range: Range Amp 0: 65.0 °C 
E12 
Temperature circuit calibration out of range: Range Amp 1: 25.5 °C 
E13 
Temperature circuit calibration out of range: Range Amp 1: 34.5 °C 
E14 
Temperature circuit calibration out of range: Range Amp 2: 34.5 °C 
E15 
Temperature circuit calibration out of range: Range Amp 2: 44.6 °C 
E16 
Temperature circuit calibration out of range: Range Amp 3: 65.0 °C 
E20 
Heater wire circuit has malfunctioned, heater can not be energized: 
-  Heater plate thermal cutout tripped (section 7.2.5) 
-  Heater wire fuse (F2) is open circuit (section 7.2.2) 
-  Heater wire triac, or heater wire relay is open circuit 
E21 
Heater wire circuit has malfunctioned, heater either can not be de-energized or 
monitoring circuit has failed : 
-  Heater wire triac has shorted 
-  Heater wire sense circuit has failed 
E23 
Heater wire voltage measurement circuit faulty 
E25 
Transistor Q17 is not turning on 
E26 
Transistor Q16 is not turning on 
E27 
The heater wire relay is short circuited 
E28 
Mains voltage measurement is uncalibrated (section 8.1.3) 
E29 
The heater plate thermistor is short circuited 
E2A 
The heater plate thermistor is open circuit 
E2C 
Heater plate element circuit has malfunctioned is not turning on: 
-  Element is open circuit 
-  Heater plate triac or driver faulty 
-  Heater plate relay is faulty 
E2D 
The heater plate element is not turning off.  Heater plate triac circuit faulty. 
E30 
Power (on/off) button stuck on 
E31 
Mute button stuck on 
E32 
Mode button stuck on 
E40 
Unit not functional tested at time of manufacture