MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal)
27
6.3
“See Manual” Error Codes
The following is an explanation of the error codes that are displayed in conjunction with the See
Manual indicator flashing. A code is not displayed if the microprocessor has stopped
functioning (see technical problems - section 6.2).
Error Description of Fault
E00 No
Manual indicator flashing. A code is not displayed if the microprocessor has stopped
functioning (see technical problems - section 6.2).
Error Description of Fault
E00 No
fault
E02
Microprocessor stack overflow
E03 RAM
fault
E04 ROM
fault
E05
EEPROM version older than ROM version. Update EEPROM. Refer to Section 7.2.6
E06
EEPROM version newer than ROM version – (old software). Contact Fisher & Paykel
Healthcare
Healthcare
E07
Model mismatch with software, contact your Fisher & Paykel Healthcare
representative.
NOTE: included in software version 7.14 onward, except 7.21.
representative.
NOTE: included in software version 7.14 onward, except 7.21.
E10
Temperature circuit calibration out of range: Range Amp 0: 25.5 °C
E11
Temperature circuit calibration out of range: Range Amp 0: 65.0 °C
E12
Temperature circuit calibration out of range: Range Amp 1: 25.5 °C
E13
Temperature circuit calibration out of range: Range Amp 1: 34.5 °C
E14
Temperature circuit calibration out of range: Range Amp 2: 34.5 °C
E15
Temperature circuit calibration out of range: Range Amp 2: 44.6 °C
E16
Temperature circuit calibration out of range: Range Amp 3: 65.0 °C
E20
Heater wire circuit has malfunctioned, heater can not be energized:
- Heater plate thermal cutout tripped (section 7.2.5)
- Heater wire fuse (F2) is open circuit (section 7.2.2)
- Heater wire triac, or heater wire relay is open circuit
- Heater plate thermal cutout tripped (section 7.2.5)
- Heater wire fuse (F2) is open circuit (section 7.2.2)
- Heater wire triac, or heater wire relay is open circuit
E21
Heater wire circuit has malfunctioned, heater either can not be de-energized or
monitoring circuit has failed :
- Heater wire triac has shorted
- Heater wire sense circuit has failed
monitoring circuit has failed :
- Heater wire triac has shorted
- Heater wire sense circuit has failed
E23
Heater wire voltage measurement circuit faulty
E25
Transistor Q17 is not turning on
E26
Transistor Q16 is not turning on
E27
The heater wire relay is short circuited
E28
Mains voltage measurement is uncalibrated (section 8.1.3)
E29
The heater plate thermistor is short circuited
E2A
The heater plate thermistor is open circuit
E2C
Heater plate element circuit has malfunctioned is not turning on:
- Element is open circuit
- Heater plate triac or driver faulty
- Heater plate relay is faulty
- Element is open circuit
- Heater plate triac or driver faulty
- Heater plate relay is faulty
E2D
The heater plate element is not turning off. Heater plate triac circuit faulty.
E30
Power (on/off) button stuck on
E31
Mute button stuck on
E32
Mode button stuck on
E40
Unit not functional tested at time of manufacture