MR850 Technical Manual   *   Revision J   * Issued November 2005   *   Ref.  185041340 (185041713 internal)
 
45
1.  Hold down the power button, then apply mains power to the humidifier.  This action places 
the humidifier into service mode. 
2.  Select service mode number 1 (Calibration Probe #1 Check) by pushing the mute button 
when '-1-' is displayed. 
3.  Insert the calibration probe with the GREY collet into the temperature / flow probe socket of 
the humidifier. 
4.  Allow the display to stabilize for a few seconds, and check the number shown on the 
humidifier display.  The table below shows the numbers displayed by the humidifier in this 
mode: 
 
TEST 
PASS 
FAIL (LOW) 
FAIL (HIGH) 
Airway Temperature 
100 
101 
102 
Chamber Temperature 
100 
104 
108 
Flow Temperature 
100 
110 
120 
Calibration Resistor 
100 
140 
180 
Overheat Heater 
Wire
LED OFF 
Heater Wire 
LED ON 
Heater Wire 
LED ON 
 
If the humidifier displays any other value than '100' or if the heater wire indicator is on, check 
that you have the correct calibration probe plugged in, otherwise the MR850 PCBs are faulty 
and they will need to be serviced or replaced (refer 7.2.3). 
 
5.  Press the power button (this causes the Calibration Probe #2 Check to be executed). 
6.  Insert the calibration probe with a BLUE collet into the temperature / flow probe socket of the 
humidifier. 
7.  Allow the displayed value to stabilize for a few seconds.  Check the humidifier display.  The 
table below shows the numbers displayed by the humidifier in this mode: 
 
TEST 
PASS 
FAIL (LOW) 
FAIL (HIGH) 
Airway Temperature 
200 
201 
202 
Chamber Temperature 
200 
204 
208 
Flow Temperature 
200 
210 
220 
Calibration Resistor 
200 
240 
280 
Overheat Heater 
Wire
LED OFF 
Heater Wire 
LED ON 
Heater Wire 
LED ON 
 
If the humidifier displays any other value than '200' or if the heater wire indicator is on, check 
that you have the correct calibration probe plugged in, otherwise the MR850 PCBs are faulty 
and they will need to be serviced or replaced (refer 7.2.3). 
 
8.  Remove the calibration probe, and press the mute button to exit service mode. 
8.1.3 
Humidifier Display Test 
This test is used to determine whether the humidifier's display is working correctly. 
 
1.  Enter the service menu (refer section 8.1.1). 
2.  Select service mode number 6 (Display test) by pushing the mute button when '-6-' is 
displayed. 
3.  Check that all of the display LEDs and indicators are turned on. 
4.  Push the mode button to change to a display cycle test.  Check that all the LEDs and 
indicators light in sequence, and that there is only ever one LED on at a time. 
5.  Push the Mute button to exit back to the service menu. 
 
If the humidifier does not pass the display test, then it is recommended that the humidifier PCBs 
are either replaced or sent for servicing (refer 7.2.3).