MR850 Technical Manual   *   Revision J   * Issued November 2005   *   Ref.  185041340 (185041713 internal)
 
58 
13 EMC 
INFORMATION 
WARNING: 
The use of accessories other than those specified by Fisher & Paykel Healthcare may 
result in increased emissions or decreased immunity
 
of the equipment
 
or system. 
 
Guidance and manufacturer’s declaration – electromagnetic emissions
 
The MR850 is intended for use in the electromagnetic environment specified below. The customer or the user of the MR850 should 
assure that it is used in such an environment. 
Emissions test 
Compliance 
Electromagnetic environment - guidance 
RF emissions CISPR 11 
Group 1 
The MR850 uses RF energy only for its internal function. 
Therefore, its RF emissions are very low and are not likely to 
cause any interference in nearby electronic equipment. 
RF emissions CISPR 11 
Class B 
Harmonic emissions 
IEC 61000-3-2 
Class B 
Voltage fluctuations/ flicker emissions 
IEC 61000-3-3 
Complies 
The MR850 is suitable for use in all establishments, including 
domestic establishments and those directly connected to the 
public low-voltage power supply network that supplies buildings 
used for domestic purposes. 
 
 
Guidance and manufacturer’s declaration – electromagnetic emissions
 
The MR850 is intended for use in the electromagnetic environment specified below. The customer or the user of the MR850 should 
assure that it is used in such an environment. 
Emissions test 
IEC 60601-1 test level 
Compliance level 
Electromagnetic environment - guidance 
Electrostatic discharge 
(ESD) 
IEC 61000-4-2 
± 6 kV contact 
± 8 kV air 
± 6 kV contact 
± 8 kV air 
Floors should be wood, concrete or ceramic 
tile. If floors are covered with synthetic 
material, the relative humidity should be at 
least 30 %. 
Electrical fast 
transient/burst 
IEC 61000-4-4 
± 2 kV for power supply 
lines 
± 1 kV for input/output lines 
± 2 kV for power supply 
lines 
± 1 kV for input/output lines 
Mains power quality should be that of a 
typical commercial or hospital environment. 
Surge 
IEC 61000-4-5 
± 1 kV differential mode 
± 2 kV common mode 
± 1 kV differential mode 
± 2 kV common mode 
Mains power quality should be that of a 
typical commercial or hospital environment. 
Surge 
IEC 61000-4-5 
± 1 kV differential mode 
± 2 kV common mode 
± 1 kV differential mode 
± 2 kV common mode 
Mains power quality should be that of a 
typical commercial or hospital environment. 
Voltage dips, short 
interruptions and voltage 
variations on power supply 
input lines 
IEC 61000-4-11 
< 5 % U
(> 95 % dip in U
T
for 0,5 cycle 
40 % U
(60 % dip in U
T
for 5 cycles 
70 % U
(30 % dip in U
T
for 25 cycles 
< 5 % U
(> 95 % dip in U
T
for 5 sec 
< 5 % U
(> 95 % dip in U
T
for 0,5 cycle 
40 % U
(60 % dip in U
T
for 5 cycles 
70 % U
(30 % dip in U
T
for 25 cycles 
< 5 % U
(> 95 % dip in U
T
for 5 sec 
Mains power quality should be that of a 
typical commercial or hospital environment. 
If the user of the MR850 requires continued 
operation during power mains interruptions, 
it is recommended that the MR850 be 
powered from an uninterruptible power 
supply or a battery. 
Power frequency (50/60 
Hz) magnetic field 
IEC 61000-4-8 
3 A/m 
3 A/m 
Power frequency magnetic fields should be 
at levels characteristic of a typical location in 
a typical commercial or hospital 
environment. 
NOTE U
is the a.c. mains voltage prior to application of the test level.