GE PT878GC Manual De Usuario

Descargar
Página de 308
April 2004
4-40
Programming Site Data
Entering the Analog Output (cont.)
2. The next prompt asks you to select the analog output type 
from a number of choices, as shown in Figure 4-19 below.
Figure 4-19:  Data Source Selection Window
a. Press 
[ENTER]
 to open the window.
b. Scroll to the desired output type.
c. Press 
[SEL]
 to confirm your selection.
d. The prompt then moves to a list of unit types. (The 
available units depend on the selection made at the Data 
Source prompt.) Scroll to the desired output unit.
e. Press 
[F3]
 (OK) to confirm your selection.
         
Select Measurement
Cancel
l
OK
Diagnostics
         Aux Input A
Volume
No Unit
Velocity
UserFunction
m/s
Pressure
Temperature
Aux Input B
Skan/Measure
Mass Flow