Programming Site Data
4-53
April  2004
Entering Reynolds Correction (cont.)
e. When you have completed entering values, press 
[F3]
 (OK) 
to confirm the table and return to the Correction Factors 
window.
You have completed entering data for a Reynolds Correction 
table.
To confirm the entries, press 
[F3]
 (OK). The meter returns to 
Operate Mode.
To leave the window without confirming the entries and return 
to Operate mode, press 
[F2]
 (Cancel) or the 
[ESC]
 key. 
Entering a Calibration Factor
The Calibration Factor is used as a reference for the flowmeter 
electronics to ensure accurate readings. To enter a Calibration 
Factor, return to the Reynolds Correction tab at the top of the 
Correction Factors option. Press the [X] arrow key to reach the 
Calibration Factor window. The window appears similar to 
Figure 4-26 below. To step through each parameter, press the [T] 
key.
Figure 4-26:  The Calibration Factor Window
Reynolds Correction
Inputs
Correction Factors
Cancel
l
OK
Calibration Factor
Edit Table
Calibration Factor
On
Off
Meter K-Factor
Data Source
K Factor   
Single
Table
10
Edit Table