Servicing the PT878GC
9-21
April 2004
Calibrating the Analog Output and Inputs
The Calibration option allows you to calibrate the analog output 
and inputs. To enter the option, scroll to the Calibrate entry on the 
Service Menu and press 
[ENTER]
. The screen appears similar to 
Figure 9-14 below. 
Figure 9-14:  The Outputs Window 
in the Calibration Option
Calibrating the Analog Output
The calibration procedure consists of calibrating the analog 
output zero point (0 or 4 mA) and then calibrating the full scale 
span point (20 mA). You enter, first the setpoint, and then the 
actual value measured with a digital ammeter or voltmeter. The 
analog outputs have a resolution of ±5.0 µA. Press the [
T
] arrow 
key to enter the Analog Output window.
 Analog Output
Calibrate/Test I/O
Analog Input
         Setpoint
Cancel
 OK
mA
4
Calibrate
          Actual  4
4 mA
20 mA
Actual 20 mA
Reset
20