GE PT878GC Manual De Usuario
April 2004
9-44
Servicing the PT878GC
Updating Software Via IrOBEX
To begin updating software via IrOBEX, scroll down to the Flash
Update entry in the Service Menu and press
Update entry in the Service Menu and press
[ENTER]
. The screen
appears similar to Figure 9-27 below.
Figure 9-27: The Flash Update Window
•
Press
[F2]
(Cancel) to return to Operate Mode without
erasing the program.
•
Press
[F3]
(OK) to confirm that you wish to erase the
program.
4. The meter asks for confirmation. Repeat the options shown in
Step 1 above. After the PT878GC reboots, the screen appears
similar to Figure 9-28 below.
similar to Figure 9-28 below.
Figure 9-28: Flash Reloading Window
Update Flash
OK
--WARNING--
This will erase the
instrumentation coding.
This will erase the
instrumentation coding.
Choose ‘Cancel’ to exit or ‘OK’
to continue.
to continue.
Cancel
GE Panametrics PCI Loader v3.0 2/26/04
[HW Rev3+]
Reload Flash via Infrared (Yes/No)?
Backup Battery: OK
Firmware CRC: Valid
IrOBEX No IrCOMM