Programming Site Data
4-5
April  2004
Parameters for Special Transducers
3. The prompt asks for the transducer frequency, to transmit a 
voltage at a rate to which the transducer can respond. 
a. From the Transducer prompt, press the [
T
] key to reach 
the Frequency prompt, and press 
[ENTER]
.
b. A drop-down list of five frequencies opens, ranging from 
0.25 to 4.00 MHz. Scroll to the frequency of your 
transducers, and press 
[ENTER]
.
4. The next prompt asks for Tw, the time delay. This parameter is 
actually the time the transducer signal spends travelling 
through the transducer and cable. The PT878GC calculates the 
flow rate by subtracting the time of the upstream and 
downstream signals, so the time delay must be subtracted out 
for an accurate measurement. GE Panametrics supplies the 
time delay on a data sheet inside the transducer case.
a. From the Frequency prompt, press the [
X
] key to reach the 
Tw prompt, and press 
[ENTER]
.
b. Use the numeric keys to enter the GE Panametrics-
supplied time delay and press 
[ENTER]
.
5. When calculating the flow rate, the PT878GC must take into 
account the wedge angle, the angle of the ultrasonic 
transmission. 
a. From the Tw prompt, press the [
T
] key to reach the 
Wedge Angle prompt, and press 
[ENTER]
.
b. Use the numeric keys to enter the GE Panametrics-
supplied wedge angle (in degrees) and press 
[ENTER]
.
6. The PT878GC must also take into account the wedge 
temperature.
a. From the Wedge Angle prompt, press the [
T
] key to reach 
the Wedge Temperature prompt, and press 
[ENTER]
.
b. Use the numeric keys to enter the wedge temperature (in 
degrees) and press 
[ENTER]
.