Chapter 3     Hardware Overview
AT-MIO/AI E Series User Manual
3-8
 National Instruments Corporation
You can program polarity and range settings on a per channel basis 
so that you can configure each input channel uniquely.
The software-programmable gain on these boards increases their 
overall flexibility by matching the input signal ranges to those that 
the ADC can accommodate.  The AT-MIO-16E-1, AT-MIO-16E-2, 
AT-MIO-64E-3, AT-MIO-16E-10, and AT-MIO-16DE-10 have 
gains of 0.5, 1, 2, 5, 10, 20, 50, and 100 and are suited for a wide 
variety of signal levels.  With the proper gain setting, you can use 
the full resolution of the ADC to measure the input signal.  
Table 3-2 shows the overall input range and precision according to 
the input range configuration and gain used.
Table 3-2.     Actual Range and Measurement Precision
Range 
Configuration
Gain
Actual Input Range
Precision
1
0 to +10 V
1.0
2.0
5.0
10.0
20.0
50.0
100.0
0 to +10 V
0 to +5 V
0 to +2 V
0 to +1 V
0 to +500 mV
0 to +200 mV
0 to +100 mV
2.44 mV
1.22 mV
488.28 
µ
V
244.14 
µ
V
122.07 
µ
V
48.83 
µ
V
24.41 
µ
V
-5 to +5 V
0.5
1.0
2.0
5.0
10.0
20.0
50.0
100.0
-10 to +10 V
-5 to +5 V
-2.5 to +2.5 V
-1 to +1 V
-500 to +500 mV
-250 to +250 mV
-100 to +100 mV
-50 to +50 mV
4.88 mV
2.44 mV
1.22 mV
488.28 
µ
V
244.14 
µ
V
122.07 
µ
V
48.83 
µ
V
24.41 
µ
V
1
The value of 1 LSB of the 12-bit ADC; that is, the voltage 
increment corresponding to a change of one count in the ADC 
12-bit count.
Note:
See Appendix A, Specifications, for absolute maximum 
ratings.