National Instruments 3.21E+04 Manual De Usuario

Descargar
Página de 181
Chapter 4     Signal Connections
 National Instruments Corporation
4-45
AT-MIO/AI E Series User Manual
UISOURCE Signal
Any PFI pin can externally input the UISOURCE signal, which is not 
available as an output on the I/O connector.  The UI counter uses the 
UISOURCE signal as a clock to time the generation of the UPDATE* 
signal.  You must configure the PFI pin you select as the source for the 
UISOURCE signal in the level-detection mode.  You can configure the 
polarity selection for the PFI pin for either active high or active low.  
Figure 4-30 shows the timing requirements for the UISOURCE signal.
Figure 4-30.  UISOURCE Signal Timing
The maximum allowed frequency is 20 MHz, with a minimum pulse 
width of 23 ns high or low.  There is no minimum frequency limitation.
Either the 20 MHz or 100 kHz internal timebase normally generates the 
UISOURCE signal unless you select some external source.
t
p
t
w
t
w
t
p
t
w
= 50 ns minimum
= 23 ns minimum