Fluke 9011 Manual De Usuario

Descargar
Página de 52
41
7 Digital Communication Interface
Interface Commands
Command Description
Command
Format
Command
Example
Returned
Returned
Example
Acceptable
Values
Read program set-point soak time
pt
pt
ti: 999
ti: 5
Set program set-point soak time to
n minutes
pt=
n
pt=5
0 to 500
Read program control mode
pc
pc
prog: {OFF or ON}
prog: OFF
Set program control mode:
pc=g[o]/s[top]/c[ont]
GO or STOP or
CONT
Start program
pc=g[o]
pc=g
Stop program
pc=s[top]
pc=s
Continue program
pc=c[ont]
pc=c
Read program function
pf
pf
pf: 9
pf: 3
Set program function to
n
pf=
n
pf=2
1 to 4
Configuration Menu
Operating Parameters Menu
Read high limit
hl
hl
hl:999
hl:126
Set high limit
hl=n
hl=90
Hot: 25 to
670°C
Cold: 25 to
140°C
Serial Interface Menu
Read serial sample setting
sa[mple]
sa
sa: 999
sa: 1
Set serial sampling setting to
n
seconds
sa[mple]=
n
sa=0
0 to 999
Read serial duplex mode (cold
side only)
du[plex]
du:
du:{FULL of HALF} du: FULL
Set serial duplex mode: (cold
side only)
du[plex]=f[ull]/h[alf]
FULL or HALF
Set serial duplex mode to full
du[plex]=f[ull]
du=f
Set serial duplex mode to half
du[plex]=h[alf]
du=h
Read serial linefeed mode (cold
side only)
lf[eed]
lf
lf: ON
Set serial linefeed mode: (cold
side only)
lf[eed]=on/of[f]
ON or OFF
Set serial linefeed mode to on
lf[eed]=on
lf=on
Set serial linefeed mode to off
lf[eed]=of[f]
lf=of
Calibration Menu (WARNING – changing the following calibration values may change the accuracy of the instrument.)
Read R0 calibration parameter
r[0]
r
r0: 999.999
r0: 100.578
Set R0 calibration parameter to
n
r[0]=n
r=100.324
95.0 to 105.0
Read ALPHA calibration
parameter
al[pha]
al
al: 9.99999999
al: 0.0038573
Controller Communications Commands continued