Chapter 4
Theory of Operation
4-8
© National Instruments Corporation
counts the total number of samples taken during a controlled operation, or 
through software control in a freerun operation.
Controlled Acquisition Mode
The PCI-1200 uses two counters, counter A0 and counter A1, to execute 
DAQ operations in controlled acquisition mode. Counter A0 counts sample 
intervals, while counter A1 counts samples. In a controlled acquisition 
mode DAQ operation, the board performs a specified number of 
conversions, and then the hardware shuts off the conversions. Counter A0 
generates the conversion pulses, and counter A1 gates off counter A0 after 
the programmed count has expired. The number of conversions in a single 
controlled acquisition mode DAQ operation is limited to a 16-bit count 
(65,535 conversions).
Freerun Acquisition Mode
The PCI-1200 uses one counter, counter A0, to execute DAQ operations in 
freerun acquisition mode. Counter A0 continuously generates the 
conversion pulses as long as GATEA0 is held at a high logic level. The 
software keeps track of the number of conversions that have occurred and 
turns off counter A0 either after the required number of conversions has 
been obtained or after some other user-defined criteria have been met. The 
number of conversions in a single freerun acquisition mode DAQ operation 
is unlimited.
Interval Scanning Acquisition Mode
The PCI-1200 uses two counters for interval scanning data acquisition. 
Counter B1 is used to time the scan interval. Counter A0 times the sample 
interval. In interval scanning analog input operations, scan sequences are 
executed at regular, specified intervals. The amount of time that elapses 
between consecutive scans within the sequence is the sample interval. The 
amount of time that elapses between consecutive scan sequences is the scan 
interval
. LabVIEW, LabWindows/CVI, other application software, and 
NI-DAQ support only multichannel interval scanning. Single-channel 
interval scanning is available only through register-level programming. 
Refer to the PCI-1200 Register-Level Programming Manual for more 
information on single-channel interval scanning.
Because interval scanning allows you to specify how frequently scan 
sequences are executed, it is useful for applications in which you need to 
sample data at regular but relatively infrequent intervals. For example, to 
sample channel 1, wait 12
µs, then sample channel 0; and if you want to