© National Instruments Corporation
5-1
5
Calibration
This chapter discusses the calibration procedures for the PCI-1200 analog 
I/O circuitry. However, the PCI-1200 is factory calibrated, and National 
Instruments can recalibrate your unit if needed. To maintain the 12-bit 
accuracy of the PCI-1200 analog input and analog output circuitry, 
recalibrate at 6-month intervals.
There are four ways to perform calibrations.
If you have LabVIEW, use the 1200 Calibrate VI. This VI is located in 
the Calibration and Configuration palette.
If you have LabWindows/CVI, use the 
Calibrate_1200
 function.
If you do not have LabVIEW or LabWindows/CVI, use the NI-DAQ 
Calibrate_1200
 function.
Use your own register-level writes to the calibration DACs and the 
EEPROM. (Use this method only if NI-DAQ does not support your 
operating system.)
To calibrate using register-level writes, you need to use the PCI-1200 
Register-Level Programmer Manual.
The PCI-1200 is software calibrated. The calibration process involves 
reading offset and gain errors from the analog input and analog output data 
areas and writing values to the appropriate calibration DACs to null the 
errors. There are four calibration DACs associated with the analog input 
circuitry and four calibration DACs associated with the analog output 
circuitry, two for each output channel. After the calibration process is 
complete, each calibration DAC is at a known value. Because these values 
are lost when the board is powered down, they are also stored in the 
onboard EEPROM for future reference. 
The factory information occupies one half of the EEPROM and is 
write-protected. The lower half of the EEPROM contains user areas for 
calibration data. There are four different user areas. 
When the PCI-1200 is powered on, or the conditions under which it is 
operating change, you must load the calibration DACs with the appropriate 
calibration constants.