eutech-instruments waterproof cyberscan pc 10 phconductivity meter Manual De Usuario

Descargar
Página de 31
Instruction Manual 
PC 10 
- 5 - 
3.2 Connecting 
the 
Probe 
The Waterproof PC 10 meter uses a special combination pH / Conductivity / Temperature 
probe.  The probe cable has a notched 6-pin connector to attach the probe to the meter.  
NOTE:  Do not substitute other probes or electrodes. For replacement probe, see 
the “Accessories” section, page 21-23.  
NOTE:  Keep connector dry and clean.  Do not touch connector with soiled hands.  
To connect the pH / Conductivity / Temperature probe: 
1.  Line up the notch and 6 pins of the meter’s connector with the holes in the probe’s 
connector.  Push down and screw the metal sleeve to lock the probe connector into 
place. See Figure 5.  
2.  To remove probe, unscrew the metal sleeve and slide up the probe connector.  While 
holding onto metal sleeve, pull probe away from the meter.  
CAUTION:  Do not pull on the probe cord or the probe wires might disconnect.  
 
 
 
 
 
Figure 5: Probe Connector