Instruction Manual 
DO 6 
 
 
19
 
4.4.2 To 
calibrate 
0% 
Saturation 
1)   Rinse the probe well with DI water or 
rinse solution. 
2)  Press the MODE key to select the % 
Saturation mode, if necessary. 
3)  Dip the probe into the zero oxygen 
solution. Wait for the reading to 
stabilize.  
4)  Press the CAL key. The display will 
show the intended percentage 
calibration point (0%) with a CAL 
indicator at the top corner for 1.5 
seconds. The CAL indicator will then 
blink and the display will show a value 
which is with respect to the factory 
default calibration. 
Note: When calibration is attempted on a 
factory calibrated absolute value 
measurement of 10.1% to 49.9%, the 
display will show the absolute value 
measurement for 1.5 seconds before 
displaying Err.1 and goes to the 
measurement mode. 
5)  Press the HOLD/ENTER key to 
confirm the calibration. The meter 
displays “CO” for 1.5 seconds and 
automatically calibrates to 0.0% 
Saturation and returns to 
measurement mode.  
 
Figure 11: 0% Saturation Calibration
 
A value with respect to the 
factory default calibration 
will be displayed 
%
CAL
A
%
HOLD
ENTER
CAL 
CAL
%
Calibration point  will be 
displayed for 1.5 seconds
%
“CO” will displayed for 
1.5 seconds
A