A new membrane can easily be fitted to the cap – see the drawing overleaf. The
membrane must be flat – if it wrinkles remove it and try again with a new one. It
is important that all parts are clean and dry. Neither O-ring nor membrane can be
used more than once.
membrane must be flat – if it wrinkles remove it and try again with a new one. It
is important that all parts are clean and dry. Neither O-ring nor membrane can be
used more than once.
4.2 Spare Parts
The Eutech probe is manufactured from tried and tested parts of the highest
quality, so the need for spare parts is minimal.
A kit can be purchased containing: A cap mounted with a membrane, an anode,
nut and 50 ml of electrolyte. Loose membranes and the tool to remove the ring
are also available.
Users with many points of measurements are recommended to stock one or more
probes. The shelf-life of a ready-to-use probe is very long – several years, and it
can be stored dry if the ambient temperature is not too high. It should be stored
upright. It takes two minutes to assemble a probe.
Do not remove the tape that protects the membrane until the probe is taken
into use.