Measurement: pH
 
 
C401/301/201/102/101 
 
 
 
22
2.6.5 Chemistry 
The monitoring and control of the pH of water is important in many 
different applications. This test method provides a quick and accurate 
method for measuring pH in the range pH5.9 - 8.2.   
A precise amount of phenol red indicator is added to a sample of water.  
The phenol red changes color from yellow at pH5.9 through to dark red at 
pH8.2. 
The amount of red color is related to its pH. The amount of 525nm light 
absorbed by the red color is accurately measured by the meter and then 
converted into the pH of the sample.