Insulet Corporation 017 Manual De Usuario

Descargar
Página de 190
121
10
The following table explains what the message could mean and 
what you should do.
Alert error message
Possible causes
Action
Meter Error 1
See user manual for 
further instructions.
Blood sample is too small.
Problem with the test strip.
Problem with the meter.
Very low blood glucose: less 
than 20 mg/dL.
1. If you have symptoms, such as weakness, sweating, nervousness, 
headache or confusion, follow the recommendations of your health-
care provider for treating hypoglycemia.
2. Conduct a control solution test using a new test strip. If the results of 
the control solution test are within the range printed on the side of 
the test strip vial, retest using blood and a new test strip.
3. If the control solution test does not work, or the error persists, call 
Customer Care.
Meter Error 2
See user manual for 
further instructions.
Problem with the test strip.
Problem with the meter.
Very high blood glucose: 
above 500 mg/dL.
HIGH control solution 
applied when temperature 
is too cold. (This applies 
only to control solution 
labeled HIGH.)
1. If you have symptoms, such as thirst, fatigue, excess urination or 
blurry vision, follow the recommendations of your healthcare pro-
vider for treating hyperglycemia.
2. Conduct a control solution test using a new test strip. If the results of 
the control solution test are within the range printed on the side of 
the test strip vial, retest using blood and a new test strip.
3. If the control solution test does not work or the error persists, call 
Customer Care.