Insulet Corporation 017 Manual De Usuario

Descargar
Página de 190
125
10
Auto-off.
Remove Pod now.
Hazard 
Alarm
Pod 
and 
PDM
Continuous tone.
Pod has been deactivated. Insulin deliv-
ery has stopped. Press OK to acknowl-
edge.
Change Pod.
Check blood glucose.
Low reservoir.
Replace Pod soon.
(This is based on the vol-
ume level you selected.)
Advisory 
Alarm
Pod 
and 
PDM
2 sets of beeps every minute for 3 
minutes.
Repeats every 60 minutes until 
acknowledged.
Escalates to Empty reservoir 
hazard alarm when empty.
Press OK to acknowledge.
Empty reservoir.
Insulin delivery stopped.
Change Pod now.
Hazard 
Alarm
Pod 
and 
PDM
Continuous tone.
This alarm occurs when the Pod's 
reservoir is out of insulin. Press OK to 
acknowledge.
Change Pod.
Check blood glucose.
Alarm
Type
Source
Beep Pattern
Action
The Low reservoir alert will 
escalate to an Empty reservoir 
hazard alarm when insulin is 
depleted. Be sure to respond to 
the alert when it first occurs.