Additional Guidance and Manufacturer’s Declaration–Electromagnetic
Emissions/Immunity
Emissions/Immunity
Electromagnetic Compatibility Precautions
Medical electrical equipment requires special precautions regarding electromagnetic
compatibility (EMC). Medical equipment must be installed and put into service according
to the EMC information provided in this document.
compatibility (EMC). Medical equipment must be installed and put into service according
to the EMC information provided in this document.
Communications Equipment
Portable and mobile radio frequency (RF) communications equipment can affect medical
electrical equipment EMC performance.
electrical equipment EMC performance.
Replacement of Cables, Accessories, or Transducers
The use of cables, accessories or transducers other than those specified in this document
with the exception of transducers or cables sold by the manufacturer of the equipment
as replacement parts for internal components, may result in increased emissions or
decreased immunity of the medical equipment.
with the exception of transducers or cables sold by the manufacturer of the equipment
as replacement parts for internal components, may result in increased emissions or
decreased immunity of the medical equipment.
Other Equipment
The System should not be used adjacent to or stacked with other equipment. If adjacent
or stacked use is necessary, the System should be observed to verify normal operation in
the configuration in which it will be used.
or stacked use is necessary, the System should be observed to verify normal operation in
the configuration in which it will be used.
Shielded Locations
The System is fully compliant with the requirements of IEC 60601-1-2 without being
located in a shielded room.
located in a shielded room.