Safety, Warnings, and Cautions
2
AB3145_en | 2014-06-09
DANGER
This equipment is not contained in a sealed cabinet. Do not use this equipment in locations where 
it can come in contact with liquids, including body fluids.
Caution
The operator must not touch or have contact simultaneously with the patient and the laser 
imaging system.
Caution
Do not use a cell phone within 2.0 m (6.6 ft) of a laser imager. This proximity includes any imager 
behind a wall adjacent to your location.
Caution
Do not use a microwave oven within 4.0 m (13.1 ft) of a laser imager. Electromagnetic radiation 
from a microwave oven is only an issue if after the oven door is closed and latched, the seal does 
not maintain an electromagnetic tight fit between the oven door and oven main housing. 
Determining if the seal has an electromagnetic tight fit requires special detection equipment.
Caution
Do not use in the presence of flammable anesthetics, oxygen, or nitrous oxide. This equipment 
does not have a gas-sealed electronics enclosure and could ignite any flammable or explosive 
gases present in its environment.
Caution
This equipment uses a DICOM network port, and is intended to connect to other medical devices 
over the network. It is not intended to be connected directly to other medical devices. Only 
qualified personnel may provide installation and service.
Caution
This device should not be used in close contact with MRI devices, due to possible very high 
magnetic fields near an MRI unit. The magnetic field in the area where this equipment is installed 
must be less than 50 G.
Caution
Do not substitute or modify any part of this equipment.