3 Care instructions
200150  rev  3
Instrumentarium Dental
21
3 Care instructions
The user must perform the following inspections monthly:
– Visually check that all visible labels are intact and legible
– Check Sigma M sensor cable(s) for wear and damages
– Inspect a test image for sensor defect
NOTE!
The sensor unit must not be opened or repaired by the user. There are
no user serviceable parts in the equipment. In case of malfunction
always contact your dealer.
3.1
C
LEANING
 
THE
 S
IGMA
 
UNIT
The Sigma M sensor unit contains sophisticated electronics. Treat the
device with care. Use a cloth moistened in mild cleaning solution
(soap) or mild disinfectant to clean the Sigma M unit.
Do not let any water or cleaning fluid to enter the device. Always follow
the instructions of the cleaning agent.
Use a new disposable hygienic cover for every sensor usage. If
required the sensor can be immersed in disinfectant solution. Follow
the disinfectant instructions carefully. However immersion should not
exceed 12 hours. Be carefull not to let any disinfectant to enter the
sensor unit.
Do not use aggressive chemicals that may damage the sensor (i.e.
hypochlorite or other bleaching chemicals). As a guideline the sensors
withstand CIDEX (glutaraldehyde 2,2 - 2,6% and inert components)
disinfecting solution.
WARNING!
DO NOT: 
- Sterilize the sensor using autoclave or an UV oven
- Immersion in bleach or alcohol content solution
- Clean the sensor using in-appropriate instruments
3 Care instructions
200150  rev  3
Instrumentarium Dental
21
3 Care instructions
The user must perform the following inspections monthly:
– Visually check that all visible labels are intact and legible
– Check Sigma M sensor cable(s) for wear and damages
– Inspect a test image for sensor defect
NOTE!
The sensor unit must not be opened or repaired by the user. There are
no user serviceable parts in the equipment. In case of malfunction
always contact your dealer.
3.1
C
LEANING
 
THE
 S
IGMA
 
UNIT
The Sigma M sensor unit contains sophisticated electronics. Treat the
device with care. Use a cloth moistened in mild cleaning solution
(soap) or mild disinfectant to clean the Sigma M unit.
Do not let any water or cleaning fluid to enter the device. Always follow
the instructions of the cleaning agent.
Use a new disposable hygienic cover for every sensor usage. If
required the sensor can be immersed in disinfectant solution. Follow
the disinfectant instructions carefully. However immersion should not
exceed 12 hours. Be carefull not to let any disinfectant to enter the
sensor unit.
Do not use aggressive chemicals that may damage the sensor (i.e.
hypochlorite or other bleaching chemicals). As a guideline the sensors
withstand CIDEX (glutaraldehyde 2,2 - 2,6% and inert components)
disinfecting solution.
WARNING!
DO NOT: 
- Sterilize the sensor using autoclave or an UV oven
- Immersion in bleach or alcohol content solution
- Clean the sensor using in-appropriate instruments