Guidelines for Trial-phase Implantation
21
Lead Placement
Note: Fluoroscopic evaluation of the lead position during this 
procedure will aid the physician in achieving an optimum pain 
coverage location, and is recommended.
1. Position, prep and drape the patient in the usual accepted manner. 
Inject a local anesthetic at the needle insertion site.
2. Insert the needle into the epidural space with the opening facing 
up using an angle of 45° or less. 
CAUTION: Use only the insertion needle provided in the
Lead Kit. Other needles may damage the lead. The stamped
number on the needle hub corresponds to the orientation of
the bevel, which must face up. Turning the bevel ventral
(down) may result in lead damage. An angle of more than
45º increases the risk of lead damage.
3. Remove the needle stylet and verify entry into the epidural space 
using the standard technique.