 OPERATION 
6
33
OPTIWAVE 1010
www.krohne.com
09/2014 - 4003537401 - MA OPTIWAVE 1010 R01 en
6.4  Device settings
6.4.1  Supervisor access rights
It is necessary to have "Supervisor" access rights to change device settings.
The device configuration is locked (write protected). To unlock the configuration and change 
device settings, you need to log in as a "Supervisor". If you do not log in as a Supervisor, you can 
only read the device configuration.
Procedure
•  (a) Double click on the "OPTIWAVE 1010" element in the project structure (Project pane) or (b) 
Right click on "OPTIWAVE 1010" element in the project structure (Project pane) and select 
"Parameter" in the drop-down list box.
•  Click on the Log In / Log Out menu and select "Supervisor".
•  Enter the password (the default password is 123412).
•  The device configuration is unlocked. Change the device settings. Refer to the handbook for 
more data on menu items and their parameters.
i
End of the procedure.
6.4.2  How to change device settings
At the bottom right of the window there are 3 buttons. These buttons are used only to update the 
computer database. If you click on "OK" or "Apply", no data is sent to the device. This function 
obeys FDT guidelines for certification of the DTM.
When you change the value of a menu item, a pencil symbol is shown adjacent to the changed 
value:
CAUTION!
If you change the device settings, the DTM saves the device configuration in the computer 
database. It does not send the changes to the device. For morea data about how to store the new 
settings to the device,.
Figure 6-4: Use "OK" or "Apply" to update the computer database