 OPERATION 
40 
OPTIWAVE 1010
www.krohne.com
09/2014 - 4003537401 - MA OPTIWAVE 1010 R01 en
6.5.2  Error handling
Description of errors and corrective actions
Figure 6-14: Error record screen in the DTM for PACTware™
Error Message
Cause
Corrective action
Failure (NE 107 status signal)
Microwave error
The device's hardware is defective.
Replace the signal converter.For more data, refer to 
Service warranty
 on page 44.
EEPROM error 
(Config.)
EEPROM error 
(HART)
Output current 
calibration not 
valid
No signal
No measuring 
value
Hardware error
Current output set 
to error
Out of specification (NE 107 status signal)
Signal weak
The signal amplitude is less than the 
average value. This can occur if the 
liquid is agitated or if there is foam in the 
tank. If this error occurs frequently, the 
device will possibly show  the "No 
measuring value" error message.
 1
If this error occurs frequently, it is possible that you 
must install a float with a radar target (if there is no 
float supplied with the device).
Signal strong
This error can occur if there is a large 
change in  signal amplitude.
 1
No corrective action is necessary.
Spectrum quality 
bad
The quality of the spectrum is poor. If 
this message is temporarily shown, this 
will not affect the performance of the 
device. If this message is continuously 
shown, the measured values can be 
incorrect. The error message 
"Measurement old" will then be shown. 
Possible causes are foam, vortices, 
turbulent product surface and internal 
tank elements.
Do a check of the device, tank and the process. Change 
the device settings. If necessary, contact the supplier.