4.  Attach new spontaneous breathing bladder. 
5.  Reattach lung to chest compression plate. 
6.  Reattach chest skin. 
Filling the spontaneous breathing air container 
 
Note: This simulator comes with a manual air pump to fill the spontaneous 
breathing air container.  This allows utilization of the system without the need for 
an external power source.  The user may choose to purchase an electrical or 
battery operated air pump for this operation.   
 
Manual Air Pump (provided with simulator) 
 
1.  Attach manual air pump hose to spontaneous breathing fill valve located at the side 
of the simulator torso.   
 
 
 
Needs picture to show this 
2.  Fill the air container by pumping the air pump.  
 
Note: Do not to exceed 10 bar (145 psi). A pressure relief valve is included in the 
air tank and will be activated at pressure above 10 bar (145 psi). This valve has a 
built-in reset mechanism. 
 
3.  Detach air pump from spontaneous breathing fill valve 
Changing simulator batteries 
 
When replacing worn out batteries, replace them with eight D-cell alkaline batteries. 
To replace batteries: 
1.  Detach chest skin. 
2.  Open the battery cover. 
3.  Remove and discard old batteries. 
4.  Replace new batteries as indicated on battery cover. 
5.  Align battery cover with grooves in battery compartment. 
6.  Reattach chest skin. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 