BIOTRONIK SE & Co. KG ICSPGH Manual De Usuario

Descargar
Página de 53
29
Data Transfer
Data Transfer
The follow-up data can be saved, sorted and exported. 
— Connection of an external PC system for data 
processing (e.g., CDM 3000 Cardiac Data Manager)
— Connection of an external printer for printing out 
all the data with the exception of real-time ECGs 
— Connection of a USB data stick
Documentation
Internal printer for the complete documentation of:
— all follow-up reports (e.g., program and test data, 
saved data)
— all real-time ECGs (ECG, IEGM, event markers)
Interrogating and Programming the Implant
The BIOTRONIK implants can communicate 
bidirectionally with the ICS 3000 via the programming 
head. As soon as the programming head is correctly 
positioned over the implant, the program data and all 
data stored in the implant can be transmitted to the 
ICS 3000. 
Depending on the implant, a large number of 
adjustable sets of parameters are available. These 
parameter sets are combined and saved in the 
program that is currently in use. The ICS 3000 detects 
obvious programming errors and requires these to be 
corrected before the program is transferred to the 
implant. 
Emergency Programs
Both the safe program as well as the emergency 
shock can be triggered at any time using the 
respective buttons. Contact with the implant is crucial. 
The current programming is then immediately turned 
off and the respective parameters are deactivated.
WARNING!
Trigger the safe program or an emergency shock only 
under the supervision of a physician.