correctly. Contact Medtronic Patient Services (see “How to contact
Medtronic”), or your doctor or clinic, to troubleshoot further or to replace the
Patient Assistant.
Table 1. Problems, possible causes, and possible solutions
Problem: You pressed the button, but the blue Searching light does not
flash.
• Possible cause and solution: You didn’t press the button hard
enough. Retry recording the symptom, this time making sure to press
the button firmly. Wait for a response. If you are unable to successfully
record a symptom after 2 attempts, write down your symptom and the
time it occurred. Then bring the note with you the next time you visit your
doctor or clinic.
• Possible cause and solution: The Patient Assistant might have been
in an environment that was too cold or hot. Write down your symptom
and the time it occurred. Then bring the note with you the next time you
visit your doctor or clinic. Meanwhile, allow the Patient Assistant to
come to room temperature.