Product Manual - RFID Reader LF60C SoliD
100
 
C h a p t e r   7
O p e r a t i o n
Parameter 23
(0x17)
Triggered read frequency
If a read/write error occurs, the triggered read frequency defines the time 
between two attempts to read or write a transponder; or the read frequency if 
there is a triggered read (no polling).
2  .. 10  from  1/10 s
Default: (5) 500 ms
Parameter 24
(0x18)
r/w max repeat
Delay time for sensor signal to start a defined action.
0 .. 255
Default: 5
Parameter 25
(0x19)
Transponder type
This parameter defines the validity of the read transponder data.
00 ...  Read and write TIRIS type
Each transponder page consists of 8 data bytes. 
The validity of the data bytes is
 
checked by a 2 byte CRC checksum.
Default: 00
Parameter 26
(0x1A)
Sensor activity
The transponder reader offers the option of deactivating the connected sensor.
0
Sensor not activated
1
Sensor activated
Default: 1
Parameter 27
(0x1B)
Watchport for presence
Enables a message to the host if a cassette/FOUP is detected on the I/O port, or if 
it is removed from the I/O port.
A sensor is required to use this option.
0
Report nothing
1
Report cassette/FOUP is removed
2
Report cassette/FOUP is detected
3
Report cassette/FOUP is detected and cassette is removed
Default: (3) Report cassette/FOUP is detected and removed