Selecting Input Resolution:
Select the input channel resolution for your configuration. Note that the sampling rate
increases as the resolution decreases.
Select the input channel resolution for your configuration. Note that the sampling rate
increases as the resolution decreases.
The table below lists all of the possible combinations of input channels and resolutions.
The maximum input bandwidth (-3dB attenuation) is fixed at 0.262 times the sample
rate.
Total
Resolution:
Channels
Low (8 Bits)
Medium (12 Bits)
High (16 Bits)
1
n/a
n/a
960 Hz
2
n/a
n/a
960 Hz
3
n/a
640 Hz
480 Hz
4
n/a
480 Hz
320 Hz
5
400 Hz
320 Hz
320 Hz
6
400 Hz
320 Hz
160 Hz
7
320 Hz
200 Hz
160 Hz
8
300 Hz
200 Hz
160 Hz